- Dimension : 18cm x 24cm or 24cm x 30cm
- Line Density : from 79LPI to 300LPI
- Ratio : from 3:1 to 6:1
- Focal Distance : 58cm ~ 70cm
브로슈어 요청
  • A Carbon Graphite-Interspaced with Carbon-fiber covered Grid which is optimal for digital mammography units. Using the sawing process from the field of semiconductors, JPI Grid-2100 can be controlled to very small measurements, which eliminates the abnormal uniformity of density such as Moire phenomenon.
  • High Transmission Grid for Digital Imaging.
  • It improves the transmission rate which reduces the Bucky Factor This ultimately reduces patient exposure.
Digital Image Test
LineRate(L/Cm)/Ratio Tp(±10%) B(±10%) K(±10%) S(±10%)
31/4 76 1.73 1.36 4.00
36/5 75 1.75 1.38 4.10
41/5 74 1.78 1.40 4.20
- X-ray Condition: a. 35kVp 50mAs, b. F/S 0.5mm, c. F.D 60cm, d. Mo Filteration.
- Measurement : Fluorescent Meter.
- This test was done based on IEC 60627 2001 (2nd), Clause 5.1.4b), 5.2.1, 5.2.3 The international test standard for mammo grids.
Tp(±10%) B(±10%) K(±10%) S(±10%)
-100L/r5 72 1.8 1.4 4.2
Tp(±10%) B(±10%) K(±10%) S(±10%)
-120L/r5 70 1.82 1.43 4.3
Static Mammo Solution : The grid can be customized for detectors.